胜利油田XX工贸有限责任公司LFH2001VI型COD自动分析仪用户手册目录一.警告·······································第1页二.简介·······································第1页三.功能特点···································第3页四.安装说明···································第4页五.试剂的配制和更换···························第6页六.操作方法···································第8页七.日常维护···································第24页八.注意事项···································第25页九.用户须知···································第25页十.常见故障及排除方法·························第26页注:若本用户手册内容有所更改,恕不另行通知
一.警告1.仪器工作用的催化剂具有强烈的腐蚀性,更换时应严格按照化学试剂操作规程,如不慎溅在皮肤上,应立即用大量清水冲洗并治疗
2.仪器内部软管及接头如有破损或液体泄漏应立即停机清理,并更换破损软管
3.仪器内部各玻璃件如有裂纹应急时更换
4.仪器在规定的维护周期内(4次/天、工作15天)应对各软管、玻璃件进行检查
5.不同废水的测量应采用实验室对比法,适当调整加热温度,否则会造成测量偏差
6.用完及时更换或添加蒸馏水、催化剂、氧化剂
特别当催化剂用完后,仪器仍工作,将造成液体加热时爆出
无蒸馏水、氧化剂数据将出现错误
7.当工作环境温度与仪器工作曲线标定时的环境温度发生较大变化时(如气温突降或