下载后可任意编辑JWFROEHLICH 试漏仪作业手册操作手册- MPS 100 / 150 / 300 / 350 -- MFL 300 / 360 -JW FROEHLICH 试漏仪压力范围: 5 bar用户: 订单编号:清单号:JWF 项目: 发货日期: 下载后可任意编辑 目录页码1.测试步骤....................................................................................................................42.设备初始启动(调试)...............................................................................................63.设备操作..................................................................................................................114.显示与编程...............................................................................................................175.测试参数的确定........................................................................................................396.设备维护..................................................................................................................397.故障处理..................................................................................................................40下载后可任意编辑1.测试步骤检测步骤试漏仪的泄漏检测分为若干个测试阶段. 根据检测参数的不同 (例如测试的流量、测量压力、工件几何形状,泄漏率等的不同),每一个阶段所要求的测试时间有所不同。MPS100, MPS150, MPS300 及 MPS350 型的试漏仪:MPS 100, MPS150, MPS300 及 MPS350 型号试漏仪采纳的是压力降法。预充气阶段 (非真空设备)有了启动信号后即开始预充气阶段。此时压力阀是打开的,对工件实施规定的压力.,充气阀保持打开状态。充气阶段或抽真空阶段测试系统的压力降低到量充气阶段开始时的测量压力。完成规定的测试时间后,充气阀将会关闭。平衡阶段为了在测量阶段获得稳定、可重复再现的检测条件,测试系统必须先进行平衡,使充气过程中产生的湍流稳定下来。测试系统在平衡阶段结束时的压力会被记录存储为参考值(设定点)。测量阶段假如被检测工件有任何气体泄漏,将会引起压力下降。在测量阶段,相对于参考值的任何压力变化被检测、记录,并以数字形式显示出来。排气阶段或通风阶段当密封夹具向后退回会更有利于工件的排气。下载后可任...